Click Click here

DHR/AR Rheometer: Setting Up Conditioning Steps

In this topic
Overview
Sample

Options

End of Test

Overview

Conditioning steps are usually added to a procedure to change settings including environmental conditions, axial force adjustment, and pre-shear options. They are typically used as the first or last step in a procedure but can also be inserted between steps to change settings. Conditioning steps are not required in most tests unless the current default conditions must be overwritten.

To select a Conditioning step, see Using Experimental Procedures for detailed instructions.

The following settings can be adjusted in the conditioning step:

Back to top

Sample

The Sample conditioning block is used to adjust various settings that can be used to condition the sample, typically before starting a test. Sample conditioning includes environmental control, delays for normal force, preshearing, or equilibration. Note that during preshear, no data are sampled.

Environmental Control Option

Select the following environmental control parameters:

Wait for Axial Force

Select Wait for axial force if would like to delay the start of the test until the axial force/normal stress has decayed below a user-defined maximum value.

Preshear Options

Select Perform preshear to homogenize or orient a sample, or to break the internal structure to remove sample loading memories. Note that no data are collected during preshear.

  1. Check Perform preshear to activate preshear.
  2. Choose between Angular velocity, Shear rate, Torque and Shear stress: Enter the desired value.
  1. Select the Duration of the preshear and enter a positive value.
  2. Select Advanced (only available when Angular velocity or Shear rate are selected). Choose from the available motor mode settings: Auto, Soft, Medium, Stiff. Auto is recommended for standard operation.

Back to top

 

Equilibration

Check the Perform equilibration box to insert a delay, which allows the sample to equilibrate either by letting the temperature equilibrate or the letting torque/force relax. Specify a Duration, in seconds.

Back to top

Options

The Options conditioning step is used to set the axial force and the auto strain adjustment.

Axial force adjustment allows the control of the axial force during a test. It is necessary to use this function when the sample changes its dimensions as a result of temperature changes or reaction processes. It is not necessary to use this function when the sample has a low viscosity and flows easily. In such cases, the axial force might squeeze the sample out of the gap when the conditions are not set correctly. The axial force can be controlled in either tension for torsion experiments, or compression when using parallel plates. Note that the axial force control parameters are different for torsion and linear testing.

Auto strain adjustment changes the strain amplitude in order to hold the stress within the operation range of the instrument. Auto strain is typically applied during tests that monitor the transition of a sample from a liquid to a solid or solid to a liquid, such as in gel point or glass transition measurements. Auto strain adjusts the strain only within the specified strain amplitude range. NOTE: Auto strain is not available for DMA geometries.

Axial Force Adjustment

Purge Gas

Purge gas only (no active cooling): Set this check box when liquid nitrogen is not used to cool the ETC. The ETC would be cooled using the purge gas alone.

Back to top

End of Test

The End of Test conditioning block sets the conditions of the instrument after the experiment finishes.

  1. Check the Set temperature to set an end of test temperature different from the test temperature.
  2. Enter a value for the new temperature.
  3. Check Set temperature system idle (available only if Axial Force Adjustment is active). This turns of the environmental system; no temperature control is active. This prevents contraction forces due to sample cooling, which could destroy the instrument.

Back to top


© 2014 TA Instruments–Waters LLC | 159 Lukens Drive | New Castle, DE 19720